Human TIMP-4 Immuno-PCR Assay Kit, BIQ-ELISA™
Product Description
Specifications
Size | 1 Plate Kit, 2 Plate Kit, 5 Plate Kit |
---|---|
Species | Human |
Accession Number | Q99727 |
Gene Id | 7079 |
Gene Symbols | TIMP4 |
Protein Name / Synonyms | Metalloproteinase inhibitor 4 (Tissue inhibitor of metalloproteinases 4) (TIMP-4) |
Quantitative/Semi-Quantitative | Quantitative |
Compatible Sample Types | Cell Culture Supernatants, Plasma, Serum |
Solid Support | 96-well Microplate |
Method Of Detection | qPCR |
Design Principle | Sandwich-based |
Estimated Lead Time | 4-6 weeks |
Shipping Type | Blue ice |
Storage | -20°C |
Risk-Free Guarantee
We offer a 100% guarantee on all ELISA kits and membrane cytokine arrays.
Learn MoreAmazon Gift Cards!
$5 Amazon gift card in every kit box purchased.
Component | Size / Qty |
---|---|
96-well V-bottom Plate | 1 plate |
96-well Plate Film | 1 plate |
96-well PCR Plate | 1 plate |
1X Assay Diluent | 1 bottle |
20X Wash Buffer Concentrate | 1 bottle |
Lyophilized Standard | 1 vial |
Antibody Coated Beads | 1 vial |
Oligonucleotide Conjugated Antibody | 1 vial |
1X Elution Buffer | 1 vial |
2X PCR Master Mix | 1 vial |
Primer Mix | 1 vial |
PCR Plate Film | 1 piece |
*The PCR plate used is a 0.2 mL, non-skirted 96-well plate (Thermo Fisher, cat. no. AB0600). Please ensure compatibility with your PCR machine prior to purchase. For additional information contact technical support (techsupport@raybiotech.com).
Additional Materials Required
- Real-time PCR instrument, Bio-Rad or QuantStudio™ recommended
- Orbital 96-well plate shaker (with ability to reach 700 rpm)
- Microplate Washer (BioTek 405 TS Washer recommended) and a compatible 96-well V-bottom plate magnetic stand (ThermoFisher AM10050 recommended)
- Precision pipettes to deliver 1 μL to 1 mL volumes
- Pipette tips (barrier tips are recommended to reduce contamination)
- Adjustable 1-25 mL pipettes for reagent preparation
- Centrifuge that can spin 96 well plates at 500 g
- Nuclease-free water or deionized water
- 1.5 mL, 2 mL, or 5 mL microcentrifuge tubes
Assay Procedure Summary
- Prepare all reagents, samples and standards as instructed.
- Add 5 μL of the bead and conjugate mix, and 5 μL of each standard and sample to each well. Incubate for 2.5 hours at room temperature while shaking.
- Add 40 μL Assay Diluent and wash the plate using a Microplate Washer.
- Add 40 μL Elution buffer to each well. Incubate for 10 minutes at room temperature.
- Prepare PCR/Primer Master Mix and pipette 15 μL into each well of the PCR plate.
- Transfer 5 μL supernatant from V-bottom plate to the PCR plate.
- Run real-time PCR.